MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US IMPLANT EXTRACTOR SLAPHAMMER; HIP INSTRUMENTS : IMPACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US IMPLANT EXTRACTOR SLAPHAMMER; HIP INSTRUMENTS : IMPACTION INSTRUMENTS Back to Search Results
Catalog Number 257005250
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the threaded portion of slap hammer sheared off and broke in top of threaded inserter.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information was received: was there any patient consequences or not? if yes, please provide details.No.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary it was reported that the threaded portion of slap hammer sheared off and broke in top of threaded inserter.Please send replacement to hospital the product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device found that the implant extractor slaphammer was broken from the tip.The broken fragment was located on the hole of the summit retaining imp.Inserter and it was able to be unthreaded.The observed condition of the device can be attributed to unintended use error due to prying back and forward the instrument during use or an unintended off-axis impactions while the device is not fully threaded.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the implant extractor slaphammer would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT EXTRACTOR SLAPHAMMER
Common Device Name
HIP INSTRUMENTS : IMPACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key20225484
Report Number1818910-2024-19395
Device Sequence Number3339041
Product Code FZY
UDI-Device Identifier10603295143239
UDI-Public(01)10603295143239(10)SO2049557
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257005250
Device Lot NumberSO2049557
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/04/2024
Supplement Date Received by Manufacturer10/17/2024
10/30/2024
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date10/18/2024
10/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-