Catalog Number 208-24-11
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Cyst(s) (1800); Pain (1994); Scar Tissue (2060); Synovitis (2094); Osteolysis (2377); Swelling/ Edema (4577); Joint effusion (4849); Aseptic loosening (4883)
Date of Event 06/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices unknown.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 144 months after a left knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, pain, swelling, effusion, cyst and osteolysis.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and fracture or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b5, h6, h11.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Event or Problem Description
Reason for revision: mechanical loosening of internal left knee prosthetic joint, initial encounter.This is 1 of 3 reports for this event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: h3, h6, h11 this is 1 of 2 reports the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.The reason for the reported revision cannot be confirmed from the information provided but may be the result of prosthesis wear and loosening or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear and loosening cannot be confirmed and potential contributions of manufacturing, user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images radiographs, or relevant clinical information were provided a review of complaint history determined that no further action is required as no trends were identified.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CC TIBIAL INSERT SZ 4, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key20225734
Report Number1038671-2024-03530
Device Sequence Number3339047
Product Code JWH
UDI-Device Identifier10885862054333
UDI-Public10885862054333
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K954208
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/04/2015
Device Catalogue Number208-24-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2023
Supplement Date Received by Manufacturer02/07/2025
11/03/2025
04/07/2026
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date02/07/2025
11/24/2025
04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/06/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-