• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number G407209
Medical Device Problem Codes Difficult to Insert (1316); Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 08/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturing ref: 3005334138-2024-00353, 3005334138-2024-00352 during an atrial fibrillation procedure, the transseptal needle with stylet went through the introducer during transseptal puncture.The introducer was exchanged, however the same issue occurred.The procedure was then cancelled due to a thrombus noted in the left atrium.The thrombus was not present prior to the procedure.The physicians considered it a medical condition of the patient.The patient's neurological status was checked, and everything was stable at the end of the procedure.Patient's information (e.G.Age, weight, gender, ethnicity, race, patient's initials) cannot be handled by abbott in absence of patient's written consent as required by the personal data protection national legislation.National legislation prevents the recording of such information.A written consent has not been obtained in this case; therefore, this information is not available.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6, h11.One brk transseptal needle/stylet assembly was received for evaluation.Functional testing with a brk needle was not performed because the complaint introducer is a swartz right sided (sr series) guiding introducer dilator, not a transseptal introducer dilator.For transseptal access with a brk needle, swartz left sided (sl series) transseptal guiding introducers are recommended.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported thrombus remains unknown.The cause of the brk needle insertion difficulties is consistent with the incorrect use of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key20227014
Report Number3008452825-2024-00488
Device Sequence Number11611268
Product Code DRC
UDI-Device Identifier05414734205184
UDI-Public(01)05414734205184(10)10265688(17)270228
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG407209
Device Catalogue NumberG407209
Device Lot Number10265688
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2024
Supplement Date Received by Manufacturer10/30/2024
Initial Report FDA Received Date09/16/2024
Supplement Report FDA Received Date11/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FAST-CATH SWARTZ¿ SR0 X2.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-