ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 6" SMALLBORE EXT SET W/6-PORT NANOCLAVE® MANIFOLD, CHECK VALVE, CLAMP, ROTATING; STOPCOCK, I.V. SET
|
Back to Search Results |
|
| Catalog Number |
AM6130 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event - 1 jan 2024 is a place holder since the date of event is unknown.D1 brand name: 6" smallbore ext set w/6-port nanoclave® manifold, check valve, clamp, rotating luer.D4 & h4 - because the lot# is unknown, the manufacture and expiration dates are unknown.E1 initial reporter phone number and email are unknown.H.6 the device has not yet been received, upon receiving the device, an investigation will be completed.When complete, then a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
An event on an unknown date involving a 6" smallbore ext set w/6-port nanoclave® manifold, check valve, clamp, rotating luer experienced leakage.It was reported that their cicu had leaking issues with the am6130 manifold set.They had issues where the tubing meets the manifold.Sample is available for evaluation.There was unknown patient involvement, unknown adverse event and unknown delay in therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
Received two (2) samples one (1) used.List #am6130, 6" smallbore ext set w/6-port nanoclave® manifold, check valve, clamp, rotating luer.One (1) used.List #am6130, 6" smallbore ext set w/6-port nanoclave® manifold, check valve, clamp, rotating luer, one (1) used.List #unknown, microclave.As received, no physical damage or other anomalies were observed leak tested both samples, sample #! had a small hole in the tubing where it mates with the manifold, tubing bond was sufficient.Probable cause, unintentional bending forces.Sample #2 noticed that the connection at the mixing valve was not secure, the male luer was not fully inserted into the female luer threads, causing the leak.Probable cause, connection not fully inserted during assembly in ensenada.Confirmed customer complaint of leaking where the tubing meets the manifold on both returned samples.D9 - date returned to mfg: 9/16/2024.A device history review (dhr) lot # review could not be conducted because no lot number(s) was/were identified.
|
| |
|