Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information provided: was surgery delayed due to the reported event? no, was procedure successfully completed? unknown, were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study, unknown.Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.What was done to treat the shard penetration? band aide was medical or surgical intervention required? no.Was there any special diagnostic test performed? none.What is the status of the surgeon injury today? healed.Was it difficult to break the ampoule (was extra force needed to break the ampoule)? no.Was the ampoule crushed repeatedly? no.Can you identify the lot number of the product that was used? tpbdtx.Is the device available to be returned for evaluation? yes.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).H3 evaluation: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the sample device was returned with the crushed ampoule and dried formulation inside the bulb.In addition, the package opened was returned along with the pen device.The filter is purple in color due to contact with the formulation.The sample reveals a piercing through which a glass shard protruded in the applicator bulb.Also, a piercing in the butyrate tube is observed.Visual inspection shows that all the components of the applicator are present and appropriately assembled.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|