• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEOBRONCHOSCOPE 2.0C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEOBRONCHOSCOPE 2.0C Back to Search Results
Model Number EB-1970UK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the distal end with ccd module/us probe broken.Based on the result, we concluded that it was caused due to the excessive force applied on the distal end with ccd module/us probe.In addition, our technician confirmed that the insertion flexible tube buckled; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
LINEAR ULTRASOUND VIDEOBRONCHOSCOPE 2.0C
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale 07645
2015712318
MDR Report Key20234675
Report Number9610877-2024-56907
Device Sequence Number8638058
Product Code EOQ
UDI-Device Identifier04961333148226
UDI-Public04961333148226
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K131946
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEB-1970UK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2024
Initial Report FDA Received Date09/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-