U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TGMR373715
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 08/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.H6: code b20 -the device remains implanted and was therefore not available for engineering evaluation by gore.According to the gore® tag® conformable thoracic stent graft with active control system instructions for use (ifu), complications associated with use of the gore® tag® thoracic endoprosthesis may include but are not limited to endoleaks and reoperation.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2024, the patient underwent an endovascular procedure to treat a dissection in zone 3 left subclavian artery utilizing a gore® tag® conformable thoracic stent graft with active control system.On (b)(6) 2024, the patient underwent a reintervention for proximal type 1a endoleak with two additional gore® tag® conformable thoracic stent graft with active control system.Reportedly, patient came in to emergency room in middle of the night and physician treated the endoleak with both proximal and distal extensions using gore® tag® conformable thoracic stent graft with active control system.Physician has no cause for endoleak and aneurysm growth is not available.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
jaskaran parhar
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key20235623
Report Number2017233-2024-05314
Device Sequence Number8043042
Product Code MIH
UDI-Device Identifier00733132646814
UDI-Public00733132646814
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/29/2026
Device Catalogue NumberTGMR373715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2024
Initial Report FDA Received Date09/17/2024
Date Device Manufactured07/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexFemale
Patient Weight67 KG
-
-