MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S/C & MOD CATH TRL 48/28; HIP INSTRUMENTS : ACETABULAR TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S/C & MOD CATH TRL 48/28; HIP INSTRUMENTS : ACETABULAR TRIALS Back to Search Results
Catalog Number 205548000
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that upon inspection the following parts were found.S/c & mod cath trl 48/28 - faded.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received: was any alleged device deficiency reported for the (620510101) impactor handle? the impactor handle had bad or damaged threads.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, ¿actis broach sz 6 - not attaching to broach handle, summit broach sz4 - not attaching to broach handle, actis broach sz8 - not attaching to broach handle, replacement suction cup - ripped, emphasys lnr imp tip 32 - cracked, s/c & mod cath trl 48/28 - faded¿.The product returned to medtech orthopaedics for evaluation.The medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis revealed that the s/c & mod cath trl 48/28 exhibits an overall worn condition consistent with normal and constant usage for along period of time.Due to the observed wear of the device, the product information was difficult to read.The observed condition of the returned device was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the rs/c & mod cath trl 48/28 would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities device history lot : a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (af0514) provided is not a valid finished goods lot#.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S/C & MOD CATH TRL 48/28
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20235668
Report Number1818910-2024-19459
Device Sequence Number18183383
Product Code HWT
UDI-Device Identifier10603295084655
UDI-Public(01)10603295084655(10)AF0514
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number205548000
Device Lot NumberAF0514
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer09/20/2024
10/07/2024
Initial Report FDA Received Date09/17/2024
Supplement Report FDA Received Date09/25/2024
10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-