| Model Number |
4.0PEF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Unintended Extubation (4564)
|
| Date of Event |
08/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during use, the tracheostomy tube was tear and cut from the tube base as shown in the attached image.The tube was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the tracheostomy tube had tear and cut from the tube base as shown in the attached image.The tube was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Correction: b5.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the tube was broken, presented a discoloration and its texture was stiff.A dimensional test was performed on inner and outer diameter of the cannula and found within specifications.It was reported that the tracheostomy tube had tear and cut from the tube base.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: capping or plugging the tube is not recommended because there may not be sufficient airway for the patient to breathe.The tube and accessories are supplied sterile and cannot be adequately re-sterilized for safe reuse and are intended for single patient use.The tube and accessories are for single use and single patient use only.Attempts to re-sterilize the tube and accessories may result in product failure and increased risk to the patient.The tube and obturator are for single patient use.Duration should not exceed twenty-nine (29) days.Covidien has not substantiated use of these devices beyond 29 days.Decisions about tracheostomy tube changes should be made by the responsible physician or designate using accepted medical techniques and judgement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the tube was broken, presented a discoloration and its texture was stiff.Functional testing found that the outer and inner diameters of the cannula were within specification.It was reported that the tracheostomy tube had tear and cut from the tube base.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: capping or plugging the tube is not recommended because there may not be sufficient airway for the patient to breathe.The tube and accessories are supplied sterile and cannot be adequately re-sterilized for safe reuse and are intended for single patient use.The tube and accessories are for single use and single patient use only attempts to re-sterilized the tube and accessories may result in product failure and increased risk to the patient.The tube and obturator are for single patient use.Duration should not exceed twenty-nine (29) days.Covidien has not substantiated use of these devices beyond 29 days.Decisions about tracheostomy tube changes should be made by the responsible physician or designate using accepted medical techniques and judgement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the tracheostomy tube had tear and cut from the tube base as shown in the attached image.The tube was replaced.Additional information received, the broken cannula was placed at the end of (b)(6) 2024.It was rotated with several cannulas, and the usage time must be less than 2 months and carefully followed the maintenance instructions.
|
| |
|