MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 4.0PEF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 08/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the tracheostomy tube was tear and cut from the tube base as shown in the attached image.The tube was replaced.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during use, the tracheostomy tube had tear and cut from the tube base as shown in the attached image.The tube was replaced.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Correction: b5.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the tube was broken, presented a discoloration and its texture was stiff.A dimensional test was performed on inner and outer diameter of the cannula and found within specifications.It was reported that the tracheostomy tube had tear and cut from the tube base.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: capping or plugging the tube is not recommended because there may not be sufficient airway for the patient to breathe.The tube and accessories are supplied sterile and cannot be adequately re-sterilized for safe reuse and are intended for single patient use.The tube and accessories are for single use and single patient use only.Attempts to re-sterilize the tube and accessories may result in product failure and increased risk to the patient.The tube and obturator are for single patient use.Duration should not exceed twenty-nine (29) days.Covidien has not substantiated use of these devices beyond 29 days.Decisions about tracheostomy tube changes should be made by the responsible physician or designate using accepted medical techniques and judgement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the tube was broken, presented a discoloration and its texture was stiff.Functional testing found that the outer and inner diameters of the cannula were within specification.It was reported that the tracheostomy tube had tear and cut from the tube base.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: capping or plugging the tube is not recommended because there may not be sufficient airway for the patient to breathe.The tube and accessories are supplied sterile and cannot be adequately re-sterilized for safe reuse and are intended for single patient use.The tube and accessories are for single use and single patient use only attempts to re-sterilized the tube and accessories may result in product failure and increased risk to the patient.The tube and obturator are for single patient use.Duration should not exceed twenty-nine (29) days.Covidien has not substantiated use of these devices beyond 29 days.Decisions about tracheostomy tube changes should be made by the responsible physician or designate using accepted medical techniques and judgement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during use, the tracheostomy tube had tear and cut from the tube base as shown in the attached image.The tube was replaced.Additional information received, the broken cannula was placed at the end of (b)(6) 2024.It was rotated with several cannulas, and the usage time must be less than 2 months and carefully followed the maintenance instructions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20236645
Report Number2936999-2024-01484
Device Sequence Number15476214
Product Code JOH
UDI-Device Identifier20884521779775
UDI-Public20884521779775
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4.0PEF
Device Catalogue Number4.0PEF
Device Lot Number23C1240JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2024
Supplement Date Received by Manufacturer11/11/2024
12/10/2024
Initial Report FDA Received Date09/17/2024
Supplement Report FDA Received Date11/13/2024
12/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-