Model Number 15
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that their device had unexpectedly lost power.In this state the device may not be able to deliver defibrillation therapy, if needed.There was no harm to the patient as a result of the reported event.
 
Additional Manufacturer Narrative
A stryker service representative performed an evaluation of the customer¿s device and was unable to duplicate the reported issue.A customer quality engineer reviewed the electronic records and was able to verify the occurrence of the reported issue.A conclusive root cause of the reported issue could not be determined.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.
 
Event or Problem Description
The customer contacted stryker to report that their device had unexpectedly lost power. in this state the device may not be able to deliver defibrillation therapy, if needed.There was no harm to the patient as a result of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key20237600
Report Number0003015876-2024-02646
Device Sequence Number8090184
Product Code MKJ
UDI-Device Identifier00883873912782
UDI-Public00883873912782
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001218
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2024
Supplement Date Received by Manufacturer01/15/2025
Initial Report FDA Received Date09/17/2024
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age39 YR
Patient SexMale
Patient Weight113 KG
Patient RaceWhite
-
-