| Model Number |
OTV-S7H-1D |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 - address 1: (b)(6).E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had it easy to be blackout and needed to be unplugged and restarted while being used with a camera adapter and camera system.The issue was found during a therapeutic prostatectomy procedure that was completed with a similar device.The camera adapter and camera system did not malfunction.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: the image frame is blue.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: both of the image abnormalities are due to a faulty charged coupled device secondary to a defect in the camera cable, resulting in a communication problem.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had it easy to be blackout and needed to be unplugged and restarted while being used with a camera adapter and camera system.The issue was found during a therapeutic prostatectomy procedure that was completed with a similar device.The camera adapter and camera system did not malfunction.There were no reports of patient harm.
|
| |
|