| Catalog Number |
4558PECN |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that after the nurse drew the arterial blood gas from the patient.When the cap was pushed up the blue rubber plug in the syringe moved, which caused the blood to flow out of the syringe.There was patient involvement and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|
Search Alerts/Recalls
|