SMITHS MEDICAL ASD, INC. PORTEX PRO-VENT ARTERIAL BLOOD SAMPLING; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION Back to Search Results
Catalog Number 4558PECN
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that after the nurse drew the arterial blood gas from the patient.When the cap was pushed up the blue rubber plug in the syringe moved, which caused the blood to flow out of the syringe.There was patient involvement and no patient harm/adverse event reported.
 
Additional Manufacturer Narrative
Investigation summary: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX PRO-VENT ARTERIAL BLOOD SAMPLING
Common Device Name
TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
10 bowman dr
keene NH 03431
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key20238606
Report Number1217052-2024-00106
Device Sequence Number8637608
Product Code JKA
UDI-Device Identifier20351688039108
UDI-Public(01)20351688039108(17)260505(10)4397912
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K952516
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/05/2026
Device Catalogue Number4558PECN
Device Lot Number4397912
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date09/17/2024
Supplement Report FDA Received Date12/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-