This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The manufacture date could not be identified since the subject device was not returned.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.However, since the reported event is a known phenomenon, it is possible to infer it's cause from the results of a similar and formal investigation.Therefore, it was decided that performing an investigation by using the device with the same structure or similar device is unnecessary.As a result of the formal investigation, the guide wire distal end found that its strength did not meet specifications, and the breakage pattern showed tearing, similar to the product in the reported event.Inspection of the fracture surface of the low-strength guide wire distal end revealed layered resin solidification and the presence of welds.Brittle fractures originated at the welds, making the guide wire distal end more susceptible to tearing.This suggests a molding defect in the guide wire distal end, indicating that it was not molded under optimal processing conditions.The event can be detected/prevented by following the instructions for use which state: ·when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 5.This may damage the distal tip.Olympus will continue to monitor field performance for this device.
|