| Model Number |
VTICM5 12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).See claim# (b)(4).For fellow eye.
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault.In a separate visit the lens was exchanged for a shorter length lens and the problem was resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This is report 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause of contribution of the reported issue is not indicative of a manufacturing related issue of product.Deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 removed statement regarding lens selection.(b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault.In a separate visit the lens was exchanged for a replacement lens and the problem was resolved.Cause of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
G1 typo correction: manufacturing site 1911 walker avenue.Claim (b)(4).
|
| |
|