| Model Number |
145-5091-150 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that two echelon microcatheters were found with damaged tips.The patient was undergoing coil embolization for treatment of an aneurysm.The accessed vessel was the femoral artery with a diameter of 3.78 mm.It was reported that coil embolization, after the echelon microcatheter was shaped, it was pushed to go upper under the guidance of the nerve guidewire and attempted to be placed into the aneurysm cavity.During the placement process, it was found that the tip of the microcatheter was obviously abnormally deformed and could not maintain its original shape.The microcatheter was then withdrawn and it was found that the tip of the microcatheter was damaged.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu). the catheter was flushed as indicated in the ifu.No patient symptoms or further complications were reported as a result of this event.Additional information was received that the patient was being treated for a narrow-necked aneurysm in the right cavernous sinus segment, 4.5mm x 3.8mm.Vessel tortuosity was normal.The cause of the catheter damage was not determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: as found condition: the echelon-10 catheters were returned for analysis within a shipping box, a plastic bio-pouch, and individual sealed pouches.The lot numbers of the returned echelon-10 catheters were not distinguishable.Damage location details: no damages were found with the echelon-10 catheter hub.No bends or kinks were found with the catheter body.The coating at the catheter¿s proximal marker was found damaged (bubbled).No damages were found with the catheter distal tip.The catheter distal tip was noted to be straight.No evidence of shaping was found with the catheter distal tip.Testing/analysis: none conclusion: based on the device analysis and reported information, the customer¿s ¿catheter kink/damage¿ report was confirmed as one of the returned echelon-10 catheter distal tip was found damaged (torn/broken) at the proximal edge of the catheter¿s distal marker catheter "b".The tip damage could have occurred due to excessive force or improper technique while navigating the catheter through the vessel, due to improper shaping techniques or excessive force during the shaping process.For catheter "a," no evidence of shaping was found, suggesting it might not have been shaped correctly or the incorrect device was returned.Regarding the damaged (bubbled) coating at the catheter¿s proximal marker on both catheters, it is possible the damage occurred during manufacturing or during the use of the product.However, the cause of the damage could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the tip was attempted to be shaped with steam.
|
| |
|