MAUDE Adverse Event Report: SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.211.020TS
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device history lot: sterile part # 04.211.020ts.Sterile lot # (b)(4).Release to warehouse date: 22.Dec.2021.Expiration date: 01.Dec.2026.Supplier: (b)(4).Manufacturing site: werk selzach logistik.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Non-sterile part # 04.211.020.Non-sterile lot # (b)(4).Manufacturing site: werk selzach logistik.Release to warehouse date: 30.Nov.2021.Supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2024, the patient underwent an orfi with the va locking screw for the clavicle.The surgery was completed successfully without any surgical delay.The locking screws in question broke off under the screw heads.Although the screw heads and the plate were removed, the screw shafts remained in the body.No further information is available.This report is for one (1) 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 20mm.This is report 3 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 20MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK SELZACH LOGISTIK
bohackerweg 5
selzach
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20241299
Report Number8030965-2024-11679
Device Sequence Number8755052
Product Code HWC
UDI-Device Identifier10886982119025
UDI-Public(01)10886982119025(10)8L91462(11)211130
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103243
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.211.020TS
Device Lot Number8L91462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/04/2024
Supplement Date Received by Manufacturer09/25/2024
Initial Report FDA Received Date09/17/2024
Supplement Report FDA Received Date09/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - PLATES: CLAVICLE.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN.; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-