| Model Number |
VICM5 12.1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault.In a separate visit the lens was exchanged for a longer length lens and the problem was resolved.It has been noted that the surgeon elected a different length lens than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.Cause of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - removed statement regarding lens selection.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault.In a separate visit the lens was exchanged for a replacement lens and the problem was resolved.Cause of the event is unknown.
|
| |
|
Search Alerts/Recalls
|