Model Number 1898200T
Medical Device Problem Codes Overheating of Device (1437); Mechanical Jam (2983); Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pre-op, m4 handpiece was working intermittent, overheated, suddenly start and stop, stalled.There was 30 minutes delay in procedure.There was no patient involvement.
 
Additional Manufacturer Narrative
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information received stating device was overheated when used more than a minute when running at 5000rpm in oscillation mode.Device did not automatically activated.Handpiece run when the foot switch pressed.
 
Additional Manufacturer Narrative
Additional review of the event and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® HANDPIECE - XPS® STRAIGHTSHOT® M4
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key20244658
Report Number1045254-2024-01518
Device Sequence Number7366061
Product Code ERL
UDI-Device Identifier00643169350687
UDI-Public00643169350687
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K041413
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1898200T
Device Catalogue Number1898200T
Device Lot Number208753817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2024
Supplement Date Received by Manufacturer09/23/2024
Initial Report FDA Received Date09/18/2024
Supplement Report FDA Received Date10/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/17/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-