• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROSA VALVE; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROSA VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV701T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 08/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: during the investigation, no significant damage of the valves was determined.Permeability test: a permeability test has shown that the prosa is permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve is operating within the accepted tolerance in the horizontal position, but not within the specified tolerances in the vertical position.An accelerated outflow of the prosa could be determined.Adjustment test: the prosa valve was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function is operational; however, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valve, deposits were found in the prosa results: based on our test results, we can determine an accelerated outflow and non-adjustability on the prosa.The deposits visible in the valve have led to the malfunctions mentioned.Deposits caused by natural substances in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a prosa valve (#fv701t) was implanted during a procedure performed on (b)(6) 2017.According to the complainant, the valve showed an over-drainage and adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2024.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 16 years weight: 53 kgs height: 170 cm gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSA VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key20245585
Report Number3004721439-0240-00285
Device Sequence Number7301656
Product Code JXG
UDI-Device Identifier04041906133023
UDI-Public4041906133023
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2022
Device Model NumberFV701T
Device Catalogue NumberFV701T
Device Lot Number20035514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 YR
Patient SexMale
Patient Weight53 KG
-
-