• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX EPIDURAL NRFIT CONNECTOR, MINIPACK, TUOHY NDL, CONTINUOUS TRAYS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX EPIDURAL NRFIT CONNECTOR, MINIPACK, TUOHY NDL, CONTINUOUS TRAYS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 24-1300-22
Medical Device Problem Codes Fracture (1260); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the needle snapped and bent while in use.The event occurred while in use with the patient.There was no patient harm, but the epidural was abandoned, and a spinal was used instead.The patient required a morphine pca by the time the procedure had finished due to the spinal anesthesia wearing off.There was no medical intervention needed.The outcome of the event is resolved.
 
Additional Manufacturer Narrative
H6: evaluation codes updated.Device evaluation: four samples were received and a photo of the complaint was provided.On the samples and photos provided by the customer, it is visible that one needle is bent, the second needle is bent and also snapped upon needle hub.Third needle is snapped in two pieces and fourth needle is bent and snapped into three pieces.Based on the samples condition and complaint event description it was believed that the needle bend/snap happened due to a mechanical force which was applied on it.No trend of confirmed similar customer complaints nor internal nonconformity was identified.Therefore, this issue is considered to be isolated incident only.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX EPIDURAL NRFIT CONNECTOR, MINIPACK, TUOHY NDL, CONTINUOUS TRAYS
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL CZECH REPUBLIC A. S
olomoucká 306
hranice 753 0 1
EZ  753 01
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key20246408
Report Number3011237704-2024-00190
Device Sequence Number7397860
Product Code CAZ
UDI-Device Identifier15019517075335
UDI-Public(01)15019517075335(17)280816(10)4425523
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K965017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number24-1300-22
Device Lot Number4425523
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2024
Supplement Date Received by Manufacturer10/17/2024
Initial Report FDA Received Date09/18/2024
Supplement Report FDA Received Date11/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MORPHINE.
Patient Age56 YR
Patient SexUnknown
Patient Weight70 KG
-
-