• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTIUM USA SUPERLINE II; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTIUM USA SUPERLINE II; DENTAL IMPLANT Back to Search Results
Model Number FXS4510C
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646)
Date of Event 04/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reported the implant was removed due to osseointegration failure, 4 months after implant placement.Bone quality around the area was type i, and oral hygiene around the implant was moderate.No significant finding was observed during the implant removal surgery.Bone augmentation material was not used in implant placement surgery.The patient will need another surgical intervention.
 
Additional Manufacturer Narrative
Please see attached summary memo.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERLINE II
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer (Section G)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer Contact
eun son
6761 katella ave
cypress, CA 90630
7142260229
MDR Report Key20247742
Report Number3005503242-2024-00113
Device Sequence Number7407172
Product Code DZE
UDI-Device IdentifierD755FXS4510C0
UDI-PublicD755FXS4510C0
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160965
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFXS4510C
Device Lot NumberD27KC068S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2024
Initial Report FDA Received Date09/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
-
-