Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional products: d1: heartware ventricular assist system ¿ battery d4: model#: 1650 / catalog#: 1650 / expiration date: 31-oct-2019 / serial#:(b)(6) d9: no h3: no h4: mfg date: 15-oct-2018 h5: no h6: the codes present in section h6 correspond to components/products that comprise the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
A supplemental report is being submitted for device evaluation.Product event summary: two (2) batteries (b)(6) were not returned for evaluation.A review of the manufacturing documentation confirmed that the associated devices met all requirements for release.A review of the available log files report revealed no anomalies within the analyzed period.As a result, the reported event could not be confirmed.Applicable risk documentation and experience with events of similar circumstances were considered; events involving batteries not charging can be attributed, but not limited, to an internal battery communication error, faulty integrated circuit, improper weld, soldering defect, out of temperature range, and/or a charge time-out of eight (8) hours.Additional products: d4: serial or lot#: (b)(6).H3: yes h6: the codes present in section h6 correspond to components/products that comprise the reported event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|