| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 1 day post implantation, the patient experienced disassociation of the glenosphere.Subsequently, the patient was revised, during which it was noted that there was still some bone around the humeral stem which prevented the full seating of the humeral tray, leading to disassociation.The central screw was also found to have disengaged.The tray, poly, and glenosphere were revised.Attempts have been made and no additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).G2, country event occurred in: united kingdom.The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
|
| |
|
Event or Problem Description
|
|
No additional information is available regarding the incident.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record could not be performed as no item or lot number was provided.Dhr history review cannot be performed without product identification.A definitive root cause cannot be determined.The reported event is unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|