| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker determined that the cause of the reported issue was due to an inoperative system pcb assembly.A replacement for this part is no longer available.The device was returned to the customer unrepaired.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device was stuck on the self test screen.As a result, defibrillation therapy may have been delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|