| Catalog Number |
07K65-29 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, and in house testing.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.An increase in complaints has been observed for lot 61281ud00, however, in-house performance testing was completed which indicates the product is performing as expected.A review of the device history record did not identify any non-conformances or deviations related to the complaint issue.Labeling was reviewed and adequately addresses the issue under review.Based on the investigation, no systemic issue or product deficiency was identified for architect free t4 reagent lot 61281ud00.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect free t4 (ft4) results for 2 patients.The following data was provided (reference range: 0.70-1.48 ng/dl): (b)(6) :patient #2: initial ft4 = 1.50 ng/dl; repeat result = 1.28 ng/dl.(b)(6) :patient #3: initial ft4 = 1.53ng/dl; repeat result = 1.22 ng/dl.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated architect free t4 (ft4) results for 2 patients.The following data was provided (reference range: 0.70-1.48 ng/dl): 8/29:patient #2: initial ft4 = 1.50 ng/dl; repeat result = 1.28 ng/dl, 8/30:patient #3: initial ft4 = 1.53ng/dl; repeat result = 1.22 ng/dl.There was no impact to patient management reported.
|
| |
|
Search Alerts/Recalls
|