Catalog Number 208-25-13
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377); Implant Pain (4561); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product associated with the reported event is within the scope of recall z-0019-2022.However, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
Legal case ¿ usa case-(b)(4) lk rev 1 is associated with this case.Case-(b)(4), lk rev 2 is associated with this case.It was reported that approximately 198 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, painful right knee replacement, osteolysis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.The serial number (b)(6) is confirmed to be a part of recall number: z-0019-2022.
 
Additional Manufacturer Narrative
Report numbers: 1038671-2024-03612, 1038671-2024-03613 this follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: e1 h10 the following sections were corrected: d10 concomitant devices unknown h6 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CC TIBIAL INSERT SZ 5, 13MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key20251493
Report Number1038671-2024-03608
Device Sequence Number8163708
Product Code JWH
UDI-Device Identifier10885862054425
UDI-Public10885862054425
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K954208
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2012
Device Catalogue Number208-25-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date09/18/2024
Supplement Report FDA Received Date11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/28/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-