|
No device was returned to nuvasive for evaluation.Further, operative notes from the procedure were not provided for review of usage/technique.Additionally, no information was received from the reporting facility whether proper sterilization technique had been followed and if sterilized sets were properly sterilized and achieved a successful chemical indication of sterility prior to the case.Information obtained identified the bacterial strain in the infection as pseudomonas aeruginosa, a bacterial species commonly found in the environment.Based on the information obtained, a root cause was unable to be determined; however, not following surgical aseptic technique or cleaning and sterilization procedures may have introduced this bacterial strain into the surgical field.Labeling review: "potential adverse events and complications.As with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection" "pre-operative warnings.Care should be used in the handling and storage of the brigade implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the brigade implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the brigade implants if there is any evidence of damage.Refer to cleaning and sterilization instructions below for all non-sterile parts." "packaging.Packages for each of the components should be intact upon receipt.Devices should be carefully examined for completeness, and for lack of damage, prior to use.Damaged packages or products should not be used, and should be returned to nuvasive.The instruments and implants of the system may be supplied as either sterile or non-sterile.All implants provided non-sterile are single use and should be sterilized per instructions provided below.Instruments provided non-sterile can be single-use or reusable.Discard single-use instruments after use.Reusable instruments should be reprocessed using instructions provided below." "cleaning and decontamination.All non-sterile instruments must first be thoroughly cleaned using the validated methods prescribed in the nuvasive cleaning and sterilization instructions (doc # (b)(4)) before sterilization and introduction into a sterile surgical field.Contaminated instruments should be wiped clean of visible soil at the point of use, prior to transfer to a central processing unit for cleaning and sterilization.The validated cleaning methods include both manual and automated cleaning.Visually inspect the instruments following performance of the cleaning instructions to ensure there is no visual contamination of the instruments prior to proceeding with sterilization.If possible, contamination is present at visual inspection, repeat the cleaning steps.Contaminated instruments should not be used and should be returned to nuvasive.Contact your local representative or nuvasive directly for any additional information related to cleaning of nuvasive surgical instruments." "sterilization.All non-sterile instruments and implants are sterilizable by steam autoclave using standard hospital practices, in addition to nuvasive¿s validated parameters.In a properly functioning and calibrated steam sterilizer, effective sterilization may be achieved using the parameters prescribed in the nuvasive cleaning and sterilization instructions (doc # (b)(4)).".
|