A non-healthcare professional reported during intraocular lens (iol) implant procedure, the surgeon injected iol and noticed a scratch on optic so quickly removed iol before it finished unfolding.The procedure was completed on same day.The patient harm was not indicated.Surgeon indicated that the issue might have been user-related (scrub technician not loading iol correctly).Surgeon was unsure, how long was the lens was in the cartridge before implanting.Technician said to the surgeon that the scrub loads the lenses once the doctor switches to cortical cleanup.Additional information has been requested.
|
|
Correction information was provided in d.4.The product was not returned.Product history records were reviewed, and the documentation indicated the product met release criteria.The account indicated the use of qualified associated products.The product investigation could not identify a root cause for the reported complaint.The account indicated the product was not available to evaluate.Information was provided that the doctor indicated the issue may have been user-related, reporting that the scrub tech did not load the iol correctly.It was reported the account was also uncertain of the amount of ophthalmic viscosurgical device (ovd) used and how long the lens remained in the cartridge before implanting.The instructions for use (ifu) instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the intraocular lens (iol) to stay in the folded condition.Failure to adhere to manufacturer's recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|