ALCON RESEARCH, LLC - HUNTINGTON CLAREON TORIC UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCW0T3
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported during intraocular lens (iol) implant procedure, the surgeon injected iol and noticed a scratch on optic so quickly removed iol before it finished unfolding.The procedure was completed on same day.The patient harm was not indicated.Surgeon indicated that the issue might have been user-related (scrub technician not loading iol correctly).Surgeon was unsure, how long was the lens was in the cartridge before implanting.Technician said to the surgeon that the scrub loads the lenses once the doctor switches to cortical cleanup.Additional information has been requested.
 
Additional Manufacturer Narrative
Correction information was provided in d.4.The product was not returned.Product history records were reviewed, and the documentation indicated the product met release criteria.The account indicated the use of qualified associated products.The product investigation could not identify a root cause for the reported complaint.The account indicated the product was not available to evaluate.Information was provided that the doctor indicated the issue may have been user-related, reporting that the scrub tech did not load the iol correctly.It was reported the account was also uncertain of the amount of ophthalmic viscosurgical device (ovd) used and how long the lens remained in the cartridge before implanting.The instructions for use (ifu) instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the intraocular lens (iol) to stay in the folded condition.Failure to adhere to manufacturer's recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20256100
Report Number1119421-2024-01848
Device Sequence Number20807576
Product Code MJP
UDI-Device Identifier00380652431897
UDI-Public(01)00380652431897(17)270328(11)230330(10)15615354(21)15615354036
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCW0T3
Device Lot Number15615354
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/21/2024
Supplement Date Received by Manufacturer09/20/2024
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date10/18/2024
Date Device Manufactured03/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II LOADING FORCEPS.; MONARCH III IOL CARTRIDGE D.; MONARCH III IOL INJECTOR.; VISCOAT.
Patient Age65 YR
Patient SexFemale
-
-