• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 7.0 X 75MM REUSABLE OBTURATOR -NS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 7.0 X 75MM REUSABLE OBTURATOR -NS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 214113
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
This is report 1 of 2 for (b)(4).It was reported that during sterile processing, it was observed that the rubber on the 7.0 x 75mm reusable obturator -ns device and 8.5 x 75mm reusable obturator -ns device was peeling and exhibited splits.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: h4: the device manufacture date is currently unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: h4: the device manufacture date was unavailable in the initial medwatch report.The device manufacture date has been updated accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Investigation summary: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the device comes in used condition as expected on reusable devices.The shaft is in good condition as well as the distal tip.Upon reviewing the handle rubber, it was found that it has a little cut at the distal end near the shaft.The device was manufactured on 2015 which is 9 years old.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the 7.0 x 75mm reusable obturator -ns would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history record review: a device history review was performed and no non-conformances were identified related to the reported condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.0 X 75MM REUSABLE OBTURATOR -NS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20256416
Report Number1221934-2024-03278
Device Sequence Number7564567
Product Code LXH
UDI-Device Identifier10886705003549
UDI-Public(01)10886705003549(10)1509006
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214113
Device Lot Number1509006
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2024
Supplement Date Received by Manufacturer09/25/2024
10/07/2024
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date09/26/2024
10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-