• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 5IN 11-0 S/A BV50-3; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 5IN 11-0 S/A BV50-3; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 2881G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(6).Date sent to the fda: 9/19/2024.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: when did the needle detach or pull off from the suture (in the package, during removal from the package, during handling prior to using on the patient, or during use on the patient)? please confirm the quantity of devices involved from lot number temclh: please confirm the quantity of devices involved from lot number tjmkcz: please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2024 and suture was used.It was reported that the suture detached from the needle.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H3 evaluation: actual sample: the product was returned to ethicon for evaluation.One open folder and one used needle were received for analysis.Product code.The visual assessment of the needle noted that the swage and attachment area were noted as expected.The barrel hole of the needle was examined under magnification, and no suture remnant was found.In addition, the suture was not returned for evaluation to determine the assignable cause.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached as to what caused the reported complaint.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Representative samples: the product was returned to ethicon for evaluation.Five unopened samples were received for analysis.Product code.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand, no anomalies were observed during the evaluation.In addition, two photos were provided and showed one open box with overwraps for product code.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 5IN 11-0 S/A BV50-3
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20256549
Report Number2210968-2024-10006
Device Sequence Number8014136
Product Code GAR
UDI-Device Identifier10705031004191
UDI-Public(01)10705031004191(17)280731(10)TJMKCZ
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2881G
Device Lot NumberTJMKCZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2024
Supplement Date Received by Manufacturer10/22/2024
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date11/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-