| Model Number |
6900PTFX |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
08/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to edwards for evaluation as it remains implanted.Attempts to retrieve additional information is in process.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 29mm 6900ptfx mitral valve implanted five (5) years, six (6) months, underwent valve-in-valve procedure due to stenosis.Patient presented with congestive heart failure and shortness of breath.Tmvr was successfully performed with a 29mm 9755rsl transcatheter valve.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 29mm 6900ptfx mitral valve implanted five (5) years, six (6) months, underwent valve-in-valve procedure due to stenosis.Patient presented with congestive heart failure and shortness of breath.Tmvr was successfully performed with a 29mm 9755rsl transcatheter valve.Patient was stable at the end of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient-related factors, including hyperlipidemia and coronary artery disease.
|
| |
|