| Model Number |
D97120F5 |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a swan ganz catheter was unable to pace before use.The issue was resolved by replacing the catheter.Information including kind of surgery/examination the catheter was used for and if the patient had cardiac conduction defect was unknown.Patient demographic information was requested but unavailable.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming when the investigation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|
Additional Manufacturer Narrative
|
|
Per follow up with customer, the issue occurred during use.It was previously reported that the event occurred during device prep.A product evaluation was completed on the returned (b)(6).The reported event of unable to pace was confirmed.Continuity testing confirmed a full open condition of the proximal circuit.Cut down of the catheter body was performed to expose the pacing lead wires.Proximal leadwire was found to be broken under the balloon.Distal circuit was continuous.No visible damage or abnormality was observed from catheter body or balloon.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.A device history record review was completed and documented that device met all specifications upon distribution.A root cause cannot be determined at this time based on evidence that the failure "pacing catheter - broken leadwire - proximal" cannot be associated to manufacturing defect.The affected final work order documentation and manufacturing were verified, and no deficiencies were found.Based on the available information there is no evidence that supports or confirms the failure mode is associated to a manufacturing/ design defect.A final continuity test is performed to the electrodes during the final inspection.The instructions for use states provides the following warnings and precautions: this catheter requires special techniques for insertion and removal.Electrode dislodgement may result from pulling the catheter out through the percutaneous sheath.Avoid forceful wiping or stretching of the catheter during testing and cleaning as not to break the electrode wire circuitry.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.
|
| |
|
Search Alerts/Recalls
|