Model Number D97120F5
Medical Device Problem Code Pacing Problem (1439)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a swan ganz catheter was unable to pace before use.The issue was resolved by replacing the catheter.Information including kind of surgery/examination the catheter was used for and if the patient had cardiac conduction defect was unknown.Patient demographic information was requested but unavailable.There were no patient complications reported.
 
Additional Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming when the investigation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Additional Manufacturer Narrative
Per follow up with customer, the issue occurred during use.It was previously reported that the event occurred during device prep.A product evaluation was completed on the returned (b)(6).The reported event of unable to pace was confirmed.Continuity testing confirmed a full open condition of the proximal circuit.Cut down of the catheter body was performed to expose the pacing lead wires.Proximal leadwire was found to be broken under the balloon.Distal circuit was continuous.No visible damage or abnormality was observed from catheter body or balloon.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.A device history record review was completed and documented that device met all specifications upon distribution.A root cause cannot be determined at this time based on evidence that the failure "pacing catheter - broken leadwire - proximal" cannot be associated to manufacturing defect.The affected final work order documentation and manufacturing were verified, and no deficiencies were found.Based on the available information there is no evidence that supports or confirms the failure mode is associated to a manufacturing/ design defect.A final continuity test is performed to the electrodes during the final inspection.The instructions for use states provides the following warnings and precautions: this catheter requires special techniques for insertion and removal.Electrode dislodgement may result from pulling the catheter out through the percutaneous sheath.Avoid forceful wiping or stretching of the catheter during testing and cleaning as not to break the electrode wire circuitry.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
9492017706
MDR Report Key20256937
Report Number2015691-2024-07228
Device Sequence Number12648731
Product Code DYG
UDI-Device Identifier00690103150421
UDI-Public(01)00690103150421(17)250906(11)230907(10)65168617
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K822723
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/06/2025
Device Model NumberD97120F5
Device Lot Number65168617
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2024
Supplement Date Received by Manufacturer11/19/2024
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-