• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 3/20/90; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 3/20/90; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 357452
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The passeo-18 peripheral balloon catheter was selected for treatment.The balloon ruptured in normal stenosis.No inflation device was used to inflate the balloon.The balloon was inflated by hand and by hand you are not able to hold pressure over 6 or 10 atm.
 
Additional Manufacturer Narrative
The returned product was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinally over a length of about 14 mm.Microscopic inspection of the balloon surface showed scratches in close vicinity of the tear which have likely been caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The root cause for the reported event is most likely related to the patients anatomy.Please note that, according to the ifu, the use of a pressure monitoring device is mandatory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 3/20/90
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key20258276
Report Number1028232-2024-04934
Device Sequence Number7484903
Product Code LIT
UDI-Device Identifier07640119558639
UDI-Public(01)07640119558639(17)2703
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number357452
Device Catalogue Number357452
Device Lot Number03240726
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2024
Supplement Date Received by Manufacturer10/18/2024
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date10/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-