AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-H1100
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, the bronchovideoscope exhibited error code e216.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Three attempts were performed to obtain additional information, but no response was received from the customer.The device was returned to olympus for inspection, and the reportable malfunction was not reproduced.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the results of the investigation, it is likely that the issue occurred due to temporary poor contact on the electrical contact of the connector and system, with an abnormality in the peripheral equipment.The event can be detected/prevented by following the instructions for use which state: "important information ¿ please read before use.Precaution: caution.Turn the video system center on only when the endoscope connector is connected to the video system center.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system.Inspection of the endoscopic image.Confirm that the wli, nbi, and rdi endoscopic images are normal.5.1 troubleshooting.If any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20258568
Report Number9610595-2024-18783
Device Sequence Number7979958
Product Code EOQ
UDI-Device Identifier04953170424229
UDI-Public04953170424229
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K222861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-H1100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2024
Supplement Date Received by Manufacturer02/06/2025
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-