Catalog Number BOPT5485
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 07/24/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implants were removed due to bone loss.
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Multiple reports are being submitted for this event.A2: patient age is not provided / unknown.E1: initial reporter¿s title is not provided / unknown.
 
Event or Problem Description
(b)(6)2025 smw submitted for investigation results.(b)(6).
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Dhr review was completed for the subject lot number 2019041197.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2019041197 for similar events and three other complaints were identified.Review completed utilizing keywords: bone loss.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 TAPERED IMPLANT 5/4 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20260193
Report Number0001038806-2024-02114
Device Sequence Number18028776
Product Code DZE
UDI-Device Identifier00844868028529
UDI-Public(01)00844868028529(17)240426(10)2019041197
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/25/2024
Device Catalogue NumberBOPT5485
Device Lot Number2019041197
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date02/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight68 KG
Patient EthnicityNon Hispanic
-
-