Model Number 15
Medical Device Problem Code Key or Button Unresponsive/not Working (4063)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer's device and was unable to duplicate the reported issue.Stryker replaced the device's main keypad assembly as a precaution.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that their device's on/off button is intermittently not working.As a result, the device may be unable to power on to deliver defibrillation therapy, if needed.There was no report of patient use associated with the reported event.
 
Additional Manufacturer Narrative
The initial medwatch report was submitted in error and is a duplicate of medwatch report # 0003015876-2024-02555.
 
Event or Problem Description
The customer contacted stryker to report that their device's on/off button is intermittently not working.As a result, the device may be unable to power on to deliver defibrillation therapy, if needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key20261502
Report Number0003015876-2024-02694
Device Sequence Number8695773
Product Code MKJ
UDI-Device Identifier00883873871829
UDI-Public00883873871829
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001944
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/22/2024
Supplement Date Received by Manufacturer01/24/2025
Initial Report FDA Received Date09/19/2024
Supplement Report FDA Received Date01/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-