• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 NAVARRE OPTI DRAIN; CHRONIC CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 NAVARRE OPTI DRAIN; CHRONIC CATHETERS Back to Search Results
Catalog Number NOD8LPT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that sometime post an ultrasound guided percutaneous drainage catheter placement procedure, the catheter connector allegedly fractured.Reportedly, the catheter was replaced.There was no reported patient injury.
 
Event or Problem Description
It was reported that sometime post an ultrasound guided percutaneous ascites drainage catheter placement, the drainage catheter connector was allegedly fractured.Reportedly, the catheter was replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported catheter connector fracture as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE OPTI DRAIN
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key20265254
Report Number3006260740-2024-05345
Device Sequence Number8261099
Product Code GBX
UDI-Device Identifier00801741055119
UDI-Public(01)00801741055119(11)230801(17)260628(10)GFHT1273
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
UNKNOWN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2026
Device Catalogue NumberNOD8LPT
Device Lot NumberGFHT1273
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2024
Supplement Date Received by Manufacturer10/09/2024
Initial Report FDA Received Date09/20/2024
Supplement Report FDA Received Date11/06/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-