| Model Number |
CCWET3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Correction information was provided in d.4.Additional information was provided in h.3., h.6.And h.11.The product was not returned.Product history records were reviewed, and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.The product has not been received to evaluate.Each lens is subjected to a 100 percentage assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the company (1.50cyl), 21.5 diopter, (1.5 extended depth of focus) lens was replaced with an unspecified atiol model.The clinical reason given for the lens exchange was a "refractive surprise / anisometropia." no additional information has been provided at this time.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient experienced blurry vision.Clinical reason for explant was refractive surprise/anisemetropia.The iol was exchanged for another model company lens 1month 4 days following the initial implant procedure.Additional information has been requested.
|
| |
|
Search Alerts/Recalls
|