• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE CRS FEMORAL RT SZ 8 CEM; KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE CRS FEMORAL RT SZ 8 CEM; KNEE FEMORAL Back to Search Results
Catalog Number 150440208
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Erythema (1840); Unspecified Infection (1930); Pain (1994); Unspecified Tissue Injury (4559)
Date of Event 09/11/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient came back in to see doctor with redness (hotness) and pain in patient's rtk.After testing they determined the knee to be septic and a revision was needed.Upon opening the joint the components appeared to be loose and there was bone loss between the cement mantle and the bone.The components were removed, the joint was thoroughly irrigated and the canals were reamed out.Afterwards new revision components were placed with antibiotic cement.No further information is available at this time.Doi: (b)(6) 2022.Dor: (b)(6) 2024.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS FEMORAL RT SZ 8 CEM
Common Device Name
KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20267193
Report Number1818910-2024-19798
Device Sequence Number7419351
Product Code JWH
UDI-Device Identifier10603295042310
UDI-Public(01)10603295042310(17)310430(10)JF0307
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number150440208
Device Lot NumberJF0307
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2024
Supplement Date Received by Manufacturer09/30/2024
Initial Report FDA Received Date09/20/2024
Supplement Report FDA Received Date10/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS RP INSRT SZ 8 12MM; ATTUNE POS FEM AUG SZ 8 4MM; ATTUNE POS FEM AUG SZ 8 4MM; ATUN FEM SLV M/L 40MM HALF POR; ATUN TIB SLV M/L 45MM HALF POR; ATUNE REV RP TIB BASE SZ 8 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-