| Model Number |
8881833215 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the needle came loose from the syringe.Additional information was received from the customer and stated that the needle was completely detached from the hub of the syringe.There was no harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
A non-conformance review was conducted for lot 24b073 and there were no non-conformances related to the reported event issued during the manufacturing of this lot.Prior to a lot¿s release, the lot must be deemed acceptable by-passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There were no devices returned for evaluation; therefore, the affected product could not be evaluated to confirm the reported issue.As such, a root cause could not be determined.A corrective and preventive action search was conducted and there are no associated capas for this reported issue.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|