EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 0.3ML 31G 8MM WHOLEUNIT 10BG 500; SYRINGE, PISTON
|
Back to Search Results |
|
| Model Number |
328438 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (component code, investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|
Event or Problem Description
|
|
Consumer reported found 2 syringes from this 10 pack - when removed the needle shield, the needle hub removed with it.Dc.Lot#: 9343326a, catalog#: 328438, date of event unknown, sample status awaiting sample.
|
| |
|