|
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device malfunction was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.The event can be detected by following the applicable chapters in the instructions for use: 9, preparation and inspection: 1, visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.2, move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.Olympus will continue to monitor field performance for this device.
|