MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3544300 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast implantation procedure with mentor memorygel breast implants 300cc and experienced rupture on the right side post-operatively, which was confirmed during a physical exam.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On sep 9, 2024, additional information was received indicating the patient underwent device removal on (b)(6) 2024.During the explantation surgery, it was discovered that both breast implants were ruptured.This report is for the right breast prosthesis.Reason for device explant and/or reoperation: rupture manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On nov 28, 2024, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the siltex hpg, 300cc breast implant was found to be ruptured.A microscopic examination was performed, and the cause of the tear could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation but is more likely to occur the longer the implant is implanted.The following may cause the implant to rupture: stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|