ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
| Model Number |
A-TCSE-DD |
| Medical Device Problem Codes |
Overheating of Device (1437); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Related manufacturing ref: (b)(4).During a supraventricular tachycardia procedure, an optical issue occurred causing a delay in procedure.The catheter was connected to the tactisys, but the pressure did not display normally, indicating that the catheter was not connected.The tactisys module and radio frequency instrument were restarted, and the catheter was reseated with no resolution.The catheter was then replaced and the pressure showed normally.However, during the procedure, the temperature was not as expected and during ablation, the temperature was 39-40 degrees celsius.During ablation, the pressure gradually increased with the radiofrequency ablation time.The catheter was replaced again and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h2, h3, h6.One bi-directional, curve d-d, tacticath sensor enabled contact force ablation catheter was received for evaluation.When the returned device was connected to the tactisys quartz unit, optical fibers 1-3 met specifications for optical properties and contact force was displayed with no error messages noted.Electrode 1 and both thermocouples met specifications during electrical testing with no open or short circuits detected.In addition, the catheter met specifications during leak and occlusion testing as well as temperature accuracy.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.Udi information (d4) and 510k (g3) are not provided within this report as this product (model a-tcse-df) is an ous configuration not available in the us and is therefore not referenced in the gudid database.
|
| |
|
Search Alerts/Recalls
|
|
|