• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TCSE-DD
Medical Device Problem Codes Overheating of Device (1437); Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
Related manufacturing ref: (b)(4).During a supraventricular tachycardia procedure, an optical issue occurred causing a delay in procedure.The catheter was connected to the tactisys, but the pressure did not display normally, indicating that the catheter was not connected.The tactisys module and radio frequency instrument were restarted, and the catheter was reseated with no resolution.The catheter was then replaced and the pressure showed normally.However, during the procedure, the temperature was not as expected and during ablation, the temperature was 39-40 degrees celsius.During ablation, the pressure gradually increased with the radiofrequency ablation time.The catheter was replaced again and the procedure was completed with no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6.One bi-directional, curve d-d, tacticath sensor enabled contact force ablation catheter was received for evaluation.When the returned device was connected to the tactisys quartz unit, optical fibers 1-3 met specifications for optical properties and contact force was displayed with no error messages noted.Electrode 1 and both thermocouples met specifications during electrical testing with no open or short circuits detected.In addition, the catheter met specifications during leak and occlusion testing as well as temperature accuracy.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.Udi information (d4) and 510k (g3) are not provided within this report as this product (model a-tcse-df) is an ous configuration not available in the us and is therefore not referenced in the gudid database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-D
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA (AFD)
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key20271620
Report Number3008452825-2024-00543
Device Sequence Number19715654
Product Code OAE
UDI-Device Identifier05415067027634
UDI-Public(01)05415067027634(10)10318595(17)250930
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Expiration Date09/30/2025
Device Model NumberA-TCSE-DD
Device Catalogue NumberA-TCSE-DD
Device Lot Number10318595
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/20/2024
Supplement Report FDA Received Date12/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1