MAUDE Adverse Event Report: ALLIED HEALTHCARE PRODUCT INC. ALLIED; AHP300
FDA Home
Medical Devices
Databases
-
510(k)
|
DeNovo
|
Registration & Listing
|
Adverse Events
|
Recalls
|
PMA
|
HDE
|
Classification
|
Standards
CFR Title 21
|
Radiation-Emitting Products
|
X-Ray Assembler
|
Medsun Reports
|
CLIA
|
TPLC
ALLIED HEALTHCARE PRODUCT INC. ALLIED; AHP300
Back to Search Results
Model Number
AHP300
Medical Device Problem Code
Inaccurate Delivery (2339)
Health Effect - Clinical Code
Hypoventilation (1916)
Date of Event
01/03/2022
Type of Reportable Event
Malfunction
Additional Manufacturer Narrative
The unit was returned for repair and was recalibrated and tested.It met all specifications after the calibration was completed.
Event or Problem Description
Low tidal volume.
Search Alerts/Recalls
New Search
|
Submit an Adverse Event Report
Brand Name
ALLIED
Common Device Name
AHP300
Manufacturer
(Section D)
ALLIED HEALTHCARE PRODUCT INC.
1720 sublette ave
st. louis MO 63110
Manufacturer
(Section G)
ALLIED HEALTHCARE PRODUCT INC.
1720 sublette ave
st. louis MO 63110
Manufacturer Contact
nicole szakats
1720 sublette ave
st. louis 63110
3147712400
MDR Report Key
20273601
Report Number
1924066-2024-00052
Device Sequence Number
3469414
Product Code
BTL
UDI-Device Identifier
00026072000014
UDI-Public
(01)00026072000014
Combination Product (Y/N)
N
Initial Reporter State
OH
Initial Reporter Country
US
PMA/510(K) Number
K132021
Number of Events Summarized
1
Summary Report (Y/N)
N
Serviced by Third Party (Y/N)
Unknown
Reporter Type
Manufacturer
Report Source
Health Professional,User Facility
Initial Reporter Occupation
Administrator/Supervisor
Remedial Action
Repair
Type of Report
Initial
Report Date
(Section B)
09/11/2024
1
Device was Involved in the Event
1
Patient was Involved in the Event
Is this an Adverse Event Report?
No
Is this a Product Problem Report?
Yes
Device Model Number
AHP300
Device Catalogue Number
AHP300
Was Device Available for Evaluation?
No
Type of Report(Section G)
Initial
Initial Date Received by Manufacturer
01/03/2023
Initial Report FDA Received Date
09/20/2024
Was Device Evaluated by Manufacturer? (Y/N)
No
Date Device Manufactured
04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device
Unknown
Patient Sequence Number
1
Patient Sex
Unknown
-
-