MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 5.0PCF
Medical Device Problem Codes Use of Device Problem (1670); Gas/Air Leak (2946)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, prior to use, the cuff of the two tracheostomy tubes had a leak.There was no patient involvement.
 
Additional Manufacturer Narrative
D10 concomitant product: 5.0pcf 5.0mm id paed med cuffed x1 (lot# 23d0460jzx) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, the first tracheostomy tube had defective valve.Preventing correct inflation and deflation.The cuff of the second device tracheostomy tube did not remain inflated shortly after being implanted in the patient.There was no reported patient outcome.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the cuff presents a cut and an abrasion thus leaking.Functionally, an inflation/deflation test was performed using syringe.2cc of air was applied to the cuff and it was observed the cuff deflated after few seconds.It was reported that there was a cut in cuff.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to tube insertion.If dysfunction is detected in any part of the inflation system, the tube should not be used.Do not overinflate the cuff.Ordinarily, cuff pressure should not exceed 25 cm of h2o.Closely monitor cuff pressure under a physician¿s guidance.Overinflation may lead to permanent tracheal damage, rupture of the cuff with subsequent deflation, or cuff distortion that may cause airway blockage.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, the cuff should be pulled back.Inflate the cuff and gently move the cuff away from the distal tip of the cannula towards the neck plate as the residual air is removed by deflation.Do not use sharp instruments such as forceps or hemostats that would damage the cuff when manipulating it.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, the tracheostomy tube had defective valve preventing correct inflation and deflation.Replacement of the tube was done.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the cuff presents a cut and an abrasion thus leaking.Functionally, an inflation/deflation test was performed using syringe.2cc of air was applied to the cuff and it was observed the cuff deflated after few seconds.It was reported that there was a cut in cuff.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to tube insertion.If dysfunction is detected in any part of the inflation system, the tube should not be used.Do not overinflate the cuff.Ordinarily, cuff pressure should not exceed 25 cm of h2o.Closely monitor cuff pressure under a physician¿s guidance.Overinflation may lead to permanent tracheal damage, rupture of the cuff with subsequent deflation, or cuff distortion that may cause airway blockage.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, the cuff should be pulled back.Inflate the cuff and gently move the cuff away from the distal tip of the cannula towards the neck plate as the residual air is removed by deflation.Do not use sharp instruments such as forceps or hemostats that would damage the cuff when manipulating it.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Correction: e1 (first name, last name).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20274207
Report Number2936999-2024-01519
Device Sequence Number19681441
Product Code JOH
UDI-Device Identifier20884521779720
UDI-Public20884521779720
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 11/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5.0PCF
Device Catalogue Number5.0PCF
Device Lot Number23H1398JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2024
Supplement Date Received by Manufacturer09/27/2024
12/18/2024
12/18/2024
11/10/2025
Initial Report FDA Received Date09/20/2024
Supplement Report FDA Received Date10/10/2024
01/02/2025
01/03/2025
11/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTES ON H11
Patient SexUnknown
-
-