| Model Number |
MMT-1780KPK |
| Medical Device Problem Codes |
Break (1069); Moisture or Humidity Problem (2986); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information has been added which was not included with the initial report.The information has been provided in section h7 and h9 with this follow up report.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic minimed that the customer experienced the insulin pump suddenly shut down after exposed the pump to moisture.It was also reported that the insulin pump had a crack on the battery compartment.The customer reported no adverse event.The event involved product(s) mmt-1780kpk.Troubleshooting was performed but the issue was not resolved.No harm requiring medical intervention was reported.Advised the customer to disconnet the pump.The product mmt-1780kpk was expected to returned for analysis.
|
| |
|
Search Alerts/Recalls
|