After further review, this was found to be a non-reportable event.The event does not meet the definition of a reportable death or serious injury as the malfunction of the device did not cause or contribute to a patient death, a life-threatening event, permanent impairment of a body function or structure, required medical or surgical intervention to preclude serious impairment of a body structure or function.Additionally, the reportable malfunctions list indicates that the device in the reported incident would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Therefore, mfr# 1217052-2024-00115-00 will be cancelled.
|