• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG MEDPOR CUSTOMIZED CRANIAL-XL; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG MEDPOR CUSTOMIZED CRANIAL-XL; IMPLANT Back to Search Results
Catalog Number 54440410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
 
Event or Problem Description
It was reported that a patient a second infection.Device was removed.
 
Additional Manufacturer Narrative
Update: h6.Due to the nature of the complaint, the observation described in the reported event could not be confirmed.Expanded investigations found no manufacturing or process-related issues contributing to the infection complaint.According to the r&d medpor expert, infections are typically caused by contamination during surgery or by later exposure.This statement, originally made during the investigation of a previous complaint, was confirmed by a stryker healthcare professional.Based on all available information, the root cause of this complaint could not be determined.The investigation revealed no indication of any systematic design, material, or manufacturing-related issues.Therefore, no corrective or preventive actions are deemed necessary at this time.The complaint has been added to the complaint trend.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDPOR CUSTOMIZED CRANIAL-XL
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
verena baumgartner-kis
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key20283805
Report Number0008010177-2024-00101
Device Sequence Number8012938
Product Code FWP
UDI-Device Identifier07613252457664
UDI-Public07613252457664
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K143173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number54440410
Device Lot Number2404101047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/26/2024
Supplement Date Received by Manufacturer08/26/2024
Initial Report FDA Received Date09/23/2024
Supplement Report FDA Received Date06/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-