Catalog Number 050-95005
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A peritoneal dialysis (pd) patient's caregiver reported that during the patient's pd treatment there was a disconnection.During fill 1 of 8 part of the tubing "popped off' from the catheter line.In a follow up, the patient's pd nurse clarified the event.The nurse said that the extension set connector came loose from the extension set tubing.The patient was able to come to the clinic and had a new extension set connected.There was no harm, intervention or adverse event.The extension set is not available to return as a sample product.
 
Event or Problem Description
A peritoneal dialysis (pd) patient's caregiver reported that during the patient's pd treatment there was a disconnection.During fill 1 of 8 part of the tubing "popped off' from the catheter line.In a follow up, the patient's pd nurse clarified the event.The nurse said that the extension set connector came loose from the extension set tubing.The patient was able to come to the clinic and had a new extension set connected.There was no harm, intervention or adverse event.The extension set is not available to return as a sample product.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for the extension sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAY SAFE/LUER LOCK CATHETER EXT. 18 IN.
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key20285233
Report Number0008030665-2024-00809
Device Sequence Number7742622
Product Code KDJ
UDI-Device Identifier00840861100798
UDI-Public00840861100798
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173593
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number050-95005
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2024
Supplement Date Received by Manufacturer10/19/2024
Initial Report FDA Received Date09/23/2024
Supplement Report FDA Received Date10/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; DELFLEX PD FLUID.; LIBERTY CYCLER SET.; LIBERTY CYCLER SET.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Patient Age59 YR
Patient SexMale
Patient Weight99 KG
-
-