| Catalog Number |
050-95005 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient's caregiver reported that during the patient's pd treatment there was a disconnection.During fill 1 of 8 part of the tubing "popped off' from the catheter line.In a follow up, the patient's pd nurse clarified the event.The nurse said that the extension set connector came loose from the extension set tubing.The patient was able to come to the clinic and had a new extension set connected.There was no harm, intervention or adverse event.The extension set is not available to return as a sample product.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient's caregiver reported that during the patient's pd treatment there was a disconnection.During fill 1 of 8 part of the tubing "popped off' from the catheter line.In a follow up, the patient's pd nurse clarified the event.The nurse said that the extension set connector came loose from the extension set tubing.The patient was able to come to the clinic and had a new extension set connected.There was no harm, intervention or adverse event.The extension set is not available to return as a sample product.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for the extension sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|